• Categories
  • Soczewka 4-lustro seria "G", z rączką (VG4HAN2)

The Symbol : VG4HAN2

No Flange, 2-in-1 Handle (Posner style) - PART # VG4HAN2 version is ideal for dynamic and indentation gonioscopy

Kolor/Wzór
2500.00
opak To the virus Vault
The quantity in the package 1 opak
Shipping within 2 dni
Shipping price The Lack Of
The Availability Of dostępny

Zamówienie telefoniczne: 22 651 82 85

No Flange, 2-in-1 Handle (Posner style) - PART # VG4HAN2

4-Mirror, All-Glass Design

  • 4-mirror design allows for comprehensive examination and treatment of the trabecular meshwork with minimal lens rotation
  • Available with a
    • large ring (28.5 mm)
    • a small ring (25.5 mm)
    • 2-position handle (right/left handed)
  • No flange version is ideal for dynamic and indentation gonioscopy
  • Flanged version provides stability for laser trabeculoplasty
  • 4 mirror, Sussman style lens with unbeatable Volk optics enables high detail examination of the angle structures 
  • Also available in a handled (Posner-style) model
Mirror Angles Image Magnification Laser Spot Size Contact Diameter
4 x 64° 1.0x 1.0x

Flange: 15 mm

No Flange: 8.4 mm

Flange, Large Ring - PART # VG4

No Flange, Large Ring - PART # VG4LNF

No Flange, Small Ring - PART # VG4SNF

Safety Information

GONIOSCOPY LENSES:

Do not use a microfiber cloth, as over time these tend to collect dirt and dust which can damage the anti reflective coating on the lens.

Never use a device that shows any sign(s) of damage.

Failure to follow the point-of-use processing steps could adversely affect further decontamination steps.

Any serious incident that has occurred in relation to the device should be reported to the manufacturer and the competent authority of the member state in which the user and/or patient is established.

A thorough, manual cleaning process is recommended. Corrosive cleaning agents (i.e. acids, alkalines, etc) are not recommended. Detergent cleaning agents with neutral ph are recommended.

Do not use detergents that contain any type of emollients.

Wiping the lens with a microfiber cloth will cause lens damage. Be sure to use only a soft, lint-free
cotton cloth.

Ensure the device is completely submerged in the disinfectant solution for the entirety of the recommended or desired soak time. Do not allow the device to become unsubmerged from the disinfectant solution. Exposure to disinfectant solutions beyond the recommended soak time, and/or exposure to higher concentrations of disinfectant solution, will result in accelerated degradation of most Volk product.

Rings may discolor when exposed to Sodium Hypochlorite or Glutaraldehyde. To avoid further degradation please follow only the disinfection procedures indicated for these products in this document. This color change is purely cosmetic and will not affect the function of the lens.

Do not brush lens portion to avoid scratching. Use a soft, cotton cloth.

To avoid surface damage to contact lenses, NEVER clean the contact elements with alcohol, peroxide or acetone.

Do not use the lens when the contacting surface(s) show(s) any signs of damage.
Do not use the lens if the mirroring surface(s) show(s) any signs of damage before or during laser usage.
Do not attempt to use the lens if, for any reason, the image is unclear or unfocused.
Do not attempt to use the lens unless an adequate type and amount of coupling fluid is present between the cornea and the contacting lens surface.
Gonio lenses apart from vslt and rapid slt (vmslt) lenses are not suitable for slt (selective laser trabeculoplasty) procedure any serious incident that has occurred in relation to the device should be reported to the manufacturer and the competent authority of the member state in which the user and/or patient is established care should be taken to avoid excessive pressure on the cornea as it may affect aqueous dynamics or cause injury.

To avoid lens surface damage never clean the classic 3 mirror lenses, mini 4 mirror lens, or slt lens contact element with alcohol, peroxide, or acetone. G-series lenses may be cleaned with these chemicals.

G-series gonio lenses are not recommended for eto sterilization due to mirror degredation.
To avoid product damage, never autoclave or boil volk gonio lenses.
To avoid product damage, never subject volk gonio lenses to sterrad sterilziation.

For use by trained personnel only.

Manufacturer details

Volk Optical INC.
7893 Enterprise Dr, Mentor
OH 44060 Mentor,
United States

+1 440-942-6161
volk@volk.com

Responsible person

Rudolf Riester GmbH
Bruckstrasse, 31
72417 Jungingen
Germany

+49747792700
pfister@riester.de

The Signature Of The
Email
Ask a question